CVE-2020-12040 – Sigma Spectrum Infusion System v’s6.x (model 35700BAX) and Baxter Spectru

Sigma Spectrum Infusion System v’s6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.

 

Source: https://cve.mitre.org/cgi-bin/cvename.cgi?name=CVE-2020-12040

Reference (s):

  • https://www.us-cert.gov/ics/advisories/icsma-20-170-04
  • URL: https://www.us-cert.gov/ics/advisories/icsma-20-170-04