Get a Pentest and security assessment of your IT network.

2021-current

CVE-2020-16224 – Patient Information Center iX (PICiX) Versions B.02, C.02, C.03, Performa

Patient Information Center iX (PICiX) Versions B.02, C.02, C.03, PerformanceBridge Focal Point Version A.01, IntelliVue patient monitors MX100, MX400-MX850, and MP2-MP90 Versions N and prior, IntelliVue X3 and X2 Versions N and prior. The software parses a formatted message or structure but does not handle or incorrectly handles a length field that is inconsistent with the actual length of the associated data, causing the application on the surveillance station to restart.

 

Source: https://cve.mitre.org/cgi-bin/cvename.cgi?name=CVE-2020-16224

Reference (s):

  • https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01
  • URL: https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01